20 x 20 mm aluminium 2020 profile for laboratory instrument stands. 0.62 kg/m, 6.2 mm groove, drilled for dowel location on request.
| Material | Aluminium |
|---|---|
| Alloy Grade | 6063-T5 |
| Temper | T5 |
| Section Size | 20 x 20 mm |
| Slot | 4 x 6.2 mm groove, 6 mm mouth, 20 mm pitch, 1 slot per face, M5 and M6 |
| Length | 6 m stock, cut to list to +/-0.5 mm |
| Tolerance | EN 755-9 / GB/T 6892, wall +/-0.15 mm; machined face flatness 0.05 mm over 300 mm |
| Surface | Mill finish / clear anodised / black anodised |
| MOQ | Made to order; small quantities available from stock for trial stands |
| Lead Time | Typically 2 to 3 weeks for mill finish, plus about a week for anodising and machining |
| Loading Port | Shenzhen, Guangzhou, Ningbo |
| Supply Ability | Extrusion capacity with in-house saw cutting, drilling and machining of datum faces |
| Payment Terms | T/T 30 per cent deposit, balance against copy of bill of lading |
| Trade Terms | FOB, CIF, CFR, EXW |
| Certifications | Mill certificate to EN 10204 3.1 on request, RoHS and REACH declarations |

Three things: it has to go back to the same position after it is taken apart, it has to stay in that position while the room changes temperature, and it has to be cleanable without trapping anything. A 20 x 20 mm aluminium 2020 profile at roughly 0.62 kg/m is a good starting point for all three because it is light, dimensionally stable, and takes a wide range of slot-mounted fittings without drilling. But the profile alone does none of it. The location scheme does.
The mistake is to treat a T-slot frame as a precision structure. It is an adjustable structure. Those are different things, and the difference matters the moment an instrument has to be removed for service and refitted without a laser alignment.
Because the slot nut can sit anywhere along the slot. When you loosen a joint and retighten it, the nut slides, the profile rotates slightly, and the frame comes back together within a few tenths of a millimetre rather than within a few hundredths. For a shelf that is irrelevant. For an optical bench it is the difference between working and not.
In practice, the cheapest repeatable scheme on a 20 series frame is a dowel pin through a drilled hole in the profile and into the adjoining part, with the slot nut only providing the clamping force. The pin carries position, the bolt carries clamp. Separating those two jobs is what makes the joint repeatable.
Aluminium expands about 23.1 micrometres per metre per kelvin. A one metre run that sees a 10 K swing moves 0.23 mm. On a stand that holds a camera at one end and a sample at the other, that is a real shift — and it is not a manufacturing error, so no amount of careful assembly removes it.
| Material | Expansion over 1 m for a 10 K rise | Relative to aluminium |
|---|---|---|
| Aluminium 6063 | 0.23 mm | 1.0 |
| Carbon steel | 0.12 mm | 0.52 |
| Stainless steel 304 | 0.17 mm | 0.74 |
| Borosilicate glass | 0.033 mm | 0.14 |
| Granite | about 0.008 mm | 0.03 |
The useful number is not the absolute expansion but the differential between the frame and whatever it carries. An aluminium frame holding a borosilicate cell sees about 0.20 mm of relative movement per metre per 10 K. If the instrument’s tolerance is tighter than that, the frame cannot be the metrology reference — the measurement has to be made on a shorter path, or compensated, or the frame has to be a material with a lower coefficient.
A common and cheap mitigation: locate the instrument at one end and let the other end float. A frame that is pinned at both ends fights its own expansion and bows. A frame pinned at one end and slotted at the other moves predictably instead.
Clear anodising, for two reasons that have nothing to do with appearance. First, an anodised surface does not shed the drawing lubricant and oxide dust that mill finish carries; a mill finish profile wiped with a cleanroom wipe will keep yielding grey residue for a long time. Second, the oxide layer is hard and does not scuff into particles when a fitting is slid along the slot during adjustment.

At the floor, not inside the frame. A 20 x 20 aluminium frame is a stiff, light, lightly damped structure, so if you mount an isolation pad between the frame and the instrument, the frame becomes a resonant mass on top of a spring and can amplify floor vibration at its own natural frequency rather than attenuating it.
The working arrangement is: isolators under the frame legs, then the instrument rigidly coupled to the frame. That way the isolated mass is the whole assembly, the natural frequency is low, and the frame does not have a second resonance of its own in the working band. Levelling feet with a rubber element do both jobs at once and are the usual choice on a laboratory stand.
Through the slot, with a cover strip over the top. The 6.2 mm groove will take a sensor cable or a 4 mm pneumatic tube comfortably, and the cover strip turns the slot back into a flat face. That is the main practical advantage of the section in a laboratory: services can be added, moved and removed without a single hole in the structure.
With a location record. Three things make the difference between a stand that comes back together and one that has to be re-aligned from scratch.
Browse 20 Series Aluminium Profiles, Strut Profiles, the 20 x 20 mm four slot section, or see the 20 x 20 mm R90 rounded corner section for the full section data.
Yes, drilled to drawing on a jig, with the hole position toleranced against a stated datum face rather than against the slot. That distinction is the whole point — a dowel hole positioned from the slot inherits the slot’s tolerance, which is looser than the hole.
Around 0.05 mm over 300 mm on a machined face of the profile. That is enough for most instrument mounting but is not a surface plate; if the instrument needs better, mount it on a plate that is itself located to the frame.
For a back-of-house stand, yes. For an area that is wiped down or classified, anodised is the practical choice because mill finish keeps releasing residue onto a wipe for a long time and scuffs into particles when fittings slide.
Using 23.1 micrometre per metre per kelvin, a 1 m frame seeing a 15 K rise moves about 0.35 mm. Whether that matters depends entirely on the differential against what it carries, not on the absolute number.
Frames are normally supplied as marked parts for assembly on site, because shipping an assembled frame costs more in volume than it saves in labour. Pre-assembly and alignment at our works is available on request for instruments that are difficult to align in the room.
If the stand replaces an existing one, send a photo of the current frame with the problem area marked. Alignment complaints are almost always visible in a photo and almost never visible in a drawing.
Tell us the tolerance the stand has to hold, the temperature range, and which joints must repeat after disassembly. We will propose a location scheme and quote the profiles with the dowel holes drilled to a stated datum.
Request a quotation